Pulsed discharge ionization detector


A pulsed discharge ionization detector (pulsed discharge detector) is a detector for gas chromatography that utilizes a stable, low powered, pulsed DC discharge in helium as an ionization source.

Eluants from the GC column, flowing counter to the flow of helium from the discharge zone, are ionized by photons from the helium discharge. Bias electrode(s) focus the resulting electrons toward the collector electrode, where they cause changes in the standing current which are quantified as the detector output.

Electron capture mode

In the electron capture mode, the PDD is a selective detector for monitoring high electron affinity compounds such as freons, chlorinated pesticides, and other halogen compounds. For this type of compound, the minimum detectable quantity (MDQ) is at the femtogram (10^{-15},\mathrm{g}) or picogram (10^{-12},\mathrm{g}) level. The PDD is similar in sensitivity and response characteristics to a conventional radioactive ECD, and can be operated at temperatures up to 400°C. For operation in this mode, He and CH4 are introduced just upstream from the column exit.

Helium photoionization mode

In the helium photoionization mode, the PDD is a universal, non-destructive, high sensitivity detector. The response to both inorganic and organic compounds is linear over a wide range. Response to fixed gases is positive (increase in standing current), with an MDQ in the low ppb range. The PDD in helium photoionization mode is an excellent replacement for flame ionization detectors in petrochemical or refinery environments, where the flame and use of hydrogen can be problematic. In addition, when the helium discharge gas is doped with a suitable noble gas, such as argon, krypton, or xenon (depending on the desired cutoff point), the PDD can function as a specific photoionization detector for selective determination of aliphatics, aromatics, amines, as well as other species.

References

References

  1. W. E. Wentworth ''et al.'', "Pulsed Discharge Helium Ionization Detector ", Chromatographia, Volume 34, September/October 1992, Numbers 5-8, Pages 219-225
  2. W. E. Wentworth ''et al.'', "Pulsed Discharge Helium Ionization Detector - Universal Detector for Inorganic and Organic Compounds at the Low Picogram Level ", ''Journal of Chromatography A'', Volume 688, Issues 1-2, 30 December 1994, Pages 135-152
  3. D. S. Forsyth, "Pulsed Discharge Detector: Theory and Applications", ''Journal of Chromatography A'', Volume 1050, Issue 1, 24 September 2004, Pages 63–68
  4. Wentworth, W. E.. (Sep 1992). "Pulsed discharge helium ionization detector". Chromatographia.
Wikipedia Source

This article was imported from Wikipedia and is available under the Creative Commons Attribution-ShareAlike 4.0 License. Content has been adapted to SurfDoc format. Original contributors can be found on the article history page.

Want to explore this topic further?

Ask Mako anything about Pulsed discharge ionization detector — get instant answers, deeper analysis, and related topics.

Research with Mako

Free with your Surf account

Content sourced from Wikipedia, available under CC BY-SA 4.0.

This content may have been generated or modified by AI. CloudSurf Software LLC is not responsible for the accuracy, completeness, or reliability of AI-generated content. Always verify important information from primary sources.

Report